| Model Number |
H74939185201510 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/02/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a balloon rupture occurred.The 100% stenosed target lesion was located in the moderately tortuous and severely calcified superficial femoral artery.A 2.0mm x 150mm x 150cm coyote balloon catheter was advanced for dilation.During the procedure, the balloon ruptured upon first inflation at 8 atmospheres for 30 seconds.The device was removed without any problem, and the procedure was completed with a different device.There were no patient complications as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
G4 - premarket / 510(k) #: k111295, k162350.
|
| |
|