H3: product analysis:evaluation determined the distal bearing is detached.The likely cause of failure couldn't able to determine.H owever, it was noted that the tube is worn, at the distal it's sharp, laser markings are partially illegible and the tube does not extend/retract is stuck.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|