• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ERBE MEDICAL. LLC ERBEFLO CLEVERCAP® HYBRID CO2 TUBING/CAP SETS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ERBE MEDICAL. LLC ERBEFLO CLEVERCAP® HYBRID CO2 TUBING/CAP SETS Back to Search Results
Catalog Number 20325-239
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/24/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the hybrid co2 tubing/cap sets for olympus® scopes & co2 leaked water out through the tip of scopes.There were no reports of any patient injuries.
 
Additional Manufacturer Narrative
Sets have not been returned for analysis.However, we received similar reports of sets leaking at the distal end of the irrigation line and down scopes.At this time, two (2) primary causes of the issue have been determined: 1.Off-label use in that erbe medical llc sets are being used with other manufacturers port connectors (note: per the notes on use for the set's, only erbe accessories (i.E., erbe port connectors), etc.Are to be used with erbe tubing/cap sets.2.A change had been made to the set's connector.Currently sets are now being produced with the previous connector to minimize/eliminate leaking.Other contributing causes of leaking at the end of scopes are also being investigated.If devices are returned and the evaluation findings are different than what has already been reported and/or any other prominent causes are determined and/or remedies are implemented to resolve leaking, a follow-up mdr will be filed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ERBEFLO CLEVERCAP® HYBRID CO2 TUBING/CAP SETS
Common Device Name
CO2 TUBING/CAP SETS
Manufacturer (Section D)
ERBE MEDICAL. LLC
405 w geneva drive
tempe AZ 85282
Manufacturer (Section G)
ERBE MEDICAL, LLC
405 w geneva drive
tempe AZ 85282
Manufacturer Contact
john tartal
2225 northwest parkway
marietta, GA 30067-8764
7709554400
MDR Report Key24924130
Report Number3020712480-2026-00013
Device Sequence Number10299155
Product Code FEQ
UDI-Device Identifier04065655000375
UDI-Public04065655000375
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K162152
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Remedial Action Recall,Notification
Type of Report Initial
Report Date (Section B) 04/20/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number20325-239
Device Lot NumberW4466273
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/24/2026
Initial Report FDA Received Date04/20/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/18/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age50 YR
Patient SexUnknown
Patient Weight68 KG
-
-