• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS ELECSYS ANTI-TSHR; THYROID AUTOANTIBODY IMMUNOLOGICAL TEST SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS ELECSYS ANTI-TSHR; THYROID AUTOANTIBODY IMMUNOLOGICAL TEST SYSTEM Back to Search Results
Catalog Number 08496609190
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code Hyperthyroidism (4924)
Date of Event 03/31/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The e411 analyzer serial number was (b)(6).Issues have been identified with the elecsys anti-tshr assay, particularly around the 1.75 iu/l medical decision point.Roche diagnostics has issued an urgent medical device correction for this issue, entitled "elecsys® anti-tshr assay ¿ deviations on cobas® e 411, 402, and 801 analyzers." this correction has been reported to fda.Roche diagnostics investigated and confirmed the customers' allegations.The root cause for the variations in the elecsys anti-tshr assay was an insufficiency of the assay's routine production process.For the platform-to-platform deviation (cobas e 411 vs.Cobas e 402/e 801), differences arose due to an insufficient alignment concept between the platforms.These variations were further influenced by updates to the routine production process for this assay.For the lot-to-lot deviation on e 411, measures taken to enhance robustness led to additional shifts.Customers were advised to immediately discontinue using and discard the affected lot numbers: 840183 and 874011 on the cobas e402/e801 analyzer 840177 on the cobas e411 analyzer.Customers were advised to switch to the new unaffected lot numbers: 908953 on the cobas e402/e801 analyzer 906872 on the cobas e411 analyzer.The cobas e601 and e602 modules are not affected and can be used with all currently available lots and with the upcoming lots without restrictions.
 
Event or Problem Description
The initial reporter complained of discrepant results for 2 patient samples tested for elecsys anti-tshr (anti-tshr) on a cobas e 411 analyzer (disk system).The initial results were generated with reagent lot 84017701, expiration date 31-may-2026.The repeat results were generated with reagent lot 906872, expiration date 28-feb-2027."patient 2" initial result was 4.23 iu/l.The repeat result was 0.825 iu/l.Patient 3 initial result was 16.6 iu/l.The repeat result was 11.01 iu/l.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS ANTI-TSHR
Common Device Name
THYROID AUTOANTIBODY IMMUNOLOGICAL TEST SYSTEM
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
natalie catton
9115 hague road
indianapolis, IN 46250
MDR Report Key24924354
Report Number1823260-2026-01499
Device Sequence Number13086048
Product Code JZO
UDI-Device Identifier07613336160305
UDI-Public07613336160305
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K193313
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 04/20/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2026
Device Catalogue Number08496609190
Device Lot Number84017701
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/31/2026
Initial Report FDA Received Date04/20/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
RES 98103
Patient Sequence Number1
Patient SexUnknown
-
-