• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT DIABETES CARE INC FREESTYLE FREEDOM LITE; BLOOD GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT DIABETES CARE INC FREESTYLE FREEDOM LITE; BLOOD GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 70914-70
Medical Device Problem Code Failure to Power Up (1476)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/23/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product has been requested back for an investigation and the customer agreed to return the device; however, at this time product has not yet been received.The most probable root causes associated with this failure mode are disconnected, faulty or damaged components, software/data corruption, or misuse.However, mitigations are in place to reduce and prevent such issues.All vital manufacturing steps are validated, monitored, and verified during manufacturing to ensure the system is in conformance with the verified design.All vital functions are monitored by the system and, when necessary, function is suspended to safeguard against inaccurate results.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.All complaints and complaint trends are investigated to determine if there is a product defect/ deficiency.If a product defect/ deficiency is identified, a risk evaluation is completed and compared to the risk management report, to ensure the risk profile has not changed.Additionally, as a part of abbott¿s post-market surveillance process, all risk evaluations with associated complaint data are reviewed annually to determine if the risk profiles have changed as compared to the product risk management reports.These monitoring processes ensure that all product risk profiles remain acceptable and have a positive benefit/ risk ratio.The useful life of freestyle freedom lite meter is 5 years.As the manufacturing date of this product is 2020, it has been in distribution beyond its useful life at the time of the complaint.Since the product exceeded its useful life, it is determined to have met specification when the product was released and through its lifespan.No further investigation activities are being performed at this time.If product is returned, a physical investigation will be performed per adc's established processes and procedures and a report will be submitted upon completion of investigation.All pertinent information available to abbott diabetes care has been submitted.
 
Event or Problem Description
The reporter contacted abbott diabetes care, alleging the meter will not turn on.This complaint is being reported because the reported issue was not resolved with troubleshooting.There was no report of death, serious injury, or mistreatment associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE FREEDOM LITE
Common Device Name
BLOOD GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE INC
1360 south loop road
alameda CA 94502 7001
Manufacturer Contact
patricia reis castro
1360 south loop road
alameda, CA 94502-7001
4157205405
MDR Report Key24925223
Report Number2954323-2026-35935
Device Sequence Number13331179
Product Code NBW
UDI-Device Identifier00699073000026
UDI-Public(01)00699073000026(10)03A163D(17)310122
Combination Product (Y/N)N
PMA/510(K) Number
K092602
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 04/20/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number70914-70
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/23/2026
Initial Report FDA Received Date04/20/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/08/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age64 YR
Patient SexUnknown
Patient Weight103 KG
-
-