As reported by an edwards lifesciences field clinical specialist, regarding a 26mm sapien 3 ultra resilia valve in a pre-existing surgical valve the pulmonic position.During valve deployment within the existing surgical valve, the 26mm sapien 3 ultra resilia valve was deployed too distal.Not able to be pulled back more proximally during deployment by the implanting physician due to an acute angle and patient anatomy.With the possible risk of embolization, the wire was kept in to help hold the valve in place and a palmaz stent was then placed in a telescoping manner inside the valve and the old surgical valve to pin the valve in place, creating one long stent complex.After the valve was secured in a safe manner with this method, a non edwards valve was then implanted successfully within the stent complex.The patient was stable for the entire duration of the procedure with no vascular injury noted.
|
|
The device was used in off-label implantation in the pulmonic position.As there are no specific ifu or training materials related to pulmonic procedures, the available training materials were reviewed only for information potentially relevant to the device use.Per the instructions for use (ifu), valve malposition requiring intervention is a known potential adverse event associated with the transcatheter valve replacement (tvr) procedure.There are multiple patient and procedural factors that alone or in combination can cause or contribute to malposition, including improper positioning before deployment, poor image intensifier angle, poor coaxial alignment of the valve and delivery system, severe septal hypertrophy, minimally or bulky/severely calcified leaflets, preserved ejection fraction, significant landing zone calcification, loss of pacing capture, rapid deployment, the release of stored tension during deployment, and movement of the delivery system by the operator.The ifu cautions that incorrect sizing of the valve may lead to paravalvular leak, migration, or embolization.The device training manuals instruct the operator on proper positioning and deployment of the valve, including all procedural and anatomical considerations.Physicians are extensively trained by edwards before they are qualified to use the sapien transcatheter heart valves (all models).Training includes patient screening, device preparation, approach, deployment, imaging, procedure-specific training manuals, and proctored procedures.The correct alignment and positioning of the device at the point of deployment is emphasized as a key factor in the placement and fixation of the device.Operators are also instructed to use fluoroscopy as the primary method of visualization for positioning and deployment.In patients with high-risk anatomical features for malposition (i.E., minimal leaflet calcification, severe septal hypertrophy), balloon valvuloplasty may indicate potential balloon movement during valve deployment.In this case, there was no allegation or indication a product malfunction contributed to this adverse event.Investigation results suggest/indicate patient factors (pre-existing surgical valve in pulmonic position, tetralogy of fallot) caused or contributed to this event.A review of edwards lifesciences risk management documentation was performed for this case.The reported event is an anticipated risk of the transcatheter heart valve procedure, additional assessment of this adverse event is not required at this time.The ifu and training manuals have been reviewed and no inadequacies have been identified with regards to warnings, contraindications, and the directions/conditions for the successful use of the device.Complaint histories for all reported events are reviewed against trending control limits monthly, and any excursions above the control limits are assessed and documented as part of this monthly review.As such, neither a product risk assessment, nor corrective or preventative actions are required at this time.
|