Product complaint#: (b)(4).Date sent to the fda: 4/20/2026.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot, and no non-conformance's were identified.H3 investigational summary: the product was returned to eth for evaluation.Visual inspection revealed that one suture in several pieces and one empty opened foil were received for analysis.Product code vcp303.Upon visual assessment of the returned sample, it was noted that the suture was broken in several segments due to the process of degradation began.This likely contributed to the detachment of the needle from the suture; however, the needle was not returned for evaluation.The foil packet was visually inspected, wrinkles and holes in the cavity were observed.The condition of the package is indicative of handling and storage issues that occurred outside of ethicon control.As part of eth quality process all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.
|