(b)(4).Date sent: 4/20/2026.B3: only event year known: 2026.D4: batch #a001586p00.Additional information was requested, and the following was obtained: did the device deliver any staples? yes.If yes, were the staples formed properly? no.If yes, was the staple line complete? unsure.Did the device cut? yes.If yes, was the cut line complete? yes but jagged.If yes, was there any issue with the cut line such as jagged, dull knife, irregular, etc? jagged.Was there any difficulty opening the device? yes, would not open.Had to use manual override.If yes, how was the device removed from the patient? manual override.If yes, did the jaws eventually open? yes.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
It was reported that during an unspecified surgical procedure, it was observed that the device locked out on a patient.The stapler did not open, the staple line was not correct.There was no patient consequence nor surgical delay reported.No additional information provided.
|