MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. ECHELON; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. ECHELON; STAPLE, IMPLANTABLE Back to Search Results
Catalog Number ECH60C
Medical Device Problem Codes Failure to Form Staple (2579); Failure to Cut (2587); Difficult to Open or Close (2921)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Date sent: 4/20/2026.B3: only event year known: 2026.D4: batch #a001586p00.Additional information was requested, and the following was obtained: did the device deliver any staples? yes.If yes, were the staples formed properly? no.If yes, was the staple line complete? unsure.Did the device cut? yes.If yes, was the cut line complete? yes but jagged.If yes, was there any issue with the cut line such as jagged, dull knife, irregular, etc? jagged.Was there any difficulty opening the device? yes, would not open.Had to use manual override.If yes, how was the device removed from the patient? manual override.If yes, did the jaws eventually open? yes.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that during an unspecified surgical procedure, it was observed that the device locked out on a patient.The stapler did not open, the staple line was not correct.There was no patient consequence nor surgical delay reported.No additional information provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ECHELON
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
orla o'mahony
475 calle c
guaynabo 00969
3035526892
MDR Report Key24928086
Report Number3005075853-2026-02983
Device Sequence Number16115336
Product Code GAG
UDI-Device Identifier10705036034971
UDI-Public(01)10705036034971(17)280930(10)A99069
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
K213633
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/20/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberECH60C
Device Lot NumberA99069
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/06/2026
Supplement Date Received by Manufacturer05/19/2026
Initial Report FDA Received Date04/20/2026
Supplement Report FDA Received Date05/31/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/16/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-