| Model Number |
TSGC0801 |
| Medical Device Problem Code |
Difficult to Remove (1528)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/26/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2026, a patient presented with grade 4 functional tricuspid regurgitation (tr) for a triclip procedure.During the procedure, a triclip was successfully implanted and the procedure was discontinued.After the procedure, it was difficult to remove the sgc from the stabilizer and needed more force to detach the sgc.After removal of sgc with stabilizer from the patient, it was tested and noted that it required more force to remove it.After few attempts, it was normal to insert and remove.The final mr was reduced.There were no adverse patient effects or clinically significant delays reported.
|
| |
|
Event or Problem Description
|
|
N/a.
|
| |
|
Additional Manufacturer Narrative
|
|
All available information was investigated, and the returned device analysis was unable to confirm the reported difficulty removing the sgc guide attach from the stabilizer.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified similar complaints reported to this lot, and this device appears to be related to a potential product quality issue.The investigation determined that the reported difficult to remove may be related to a potential product quality issue.This complaint falls within the scope of an exception, as the complaint description and device code match the specific issue described in the exception.The issue is being addressed per internal operating procedures.Abbott will continue to trend the performance of these devices.
|
| |
|
Search Alerts/Recalls
|