• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. NAVLOCK; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. NAVLOCK; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number 9734590
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/25/2026
Type of Reportable Event Malfunction
Event or Problem Description
Medtronic received information regarding a navigation system being used prior to a procedure.It was reported that the post of the grey tracker broke during setup. there was no patient involvement.
 
Additional Manufacturer Narrative
H3, h6: analysis was performed for product: 9734590, lot number: 0010969067.It was reported that as reported, the post for marker #1 had been broken off and was missing.Otherwise, with the remaining markers attached and fully seated, the tracker displayed good geometry and divot error readings with normal tracking and verification.The tracker also received known good instruments without issue.Codes b01, c07 and d02 are applicable to this analysis.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NAVLOCK
Common Device Name
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC.
200 medtronic drive
lafayette CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC.
200 medtronic drive
lafayette CO 80027
Manufacturer Contact
justin ellis
8200 coral sea street ne mvs21
mounds view, MN 55112
7635265677
MDR Report Key24933075
Report Number1723170-2026-00630
Device Sequence Number14791520
Product Code OLO
UDI-Device Identifier00763000300401
UDI-Public(01)00763000300401(11)211208(10)0010969067
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K124004
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 04/20/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number9734590
Device Catalogue Number9734590
Device Lot Number0010969067
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/10/2026
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/15/2026
Initial Report FDA Received Date04/20/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/08/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-