• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION AMS 700 CX MS PUMP; PROSTHESIS, PENIS, INFLATABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION AMS 700 CX MS PUMP; PROSTHESIS, PENIS, INFLATABLE Back to Search Results
Model Number 72404231
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Codes Discomfort (2330); Unspecified Tissue Injury (4559)
Date of Event 03/27/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that this patient with an inflatable penile prosthesis (ipp) implant device experienced discomfort.A surgical procedure was performed during which the entire device was explanted and replaced.Upon explant, the physician identified one of the cylinders herniated out of the corporal body.There were no further patient complications reported.
 
Additional Manufacturer Narrative
Model number/catalog number: 72404155.Serial number: n/a.Batch/lot number: 1100311332.Model/catalog description: reservoir 65 ml pc/iz.Unique identifier (udi) #: (b)(4).The device is not available for analysis; therefore, no physical or visual analysis of the product could be performed.Review of the instructions for use (ifu) confirmed there was relevant content and sufficient guidance with respect to the circumstances described within this complaint.No updates are required to the ifu as a result of this event.A risk review was completed and confirmed that the events of discomfort, tissue damage, and device migration were defined in the product risk documentation.This event type has been accounted for during analysis to support acceptable risk benefit for the product.Based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of known inherent risk.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMS 700 CX MS PUMP
Common Device Name
PROSTHESIS, PENIS, INFLATABLE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
5086521217
MDR Report Key24933371
Report Number2124215-2026-21385
Device Sequence Number14744606
Product Code JCW
UDI-Device Identifier00878953003443
UDI-Public(01)00878953003443(17)241114(10)1100216286
Combination Product (Y/N)Y
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
N970012
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2026
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/20/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date11/14/2024
Device Model Number72404231
Device Catalogue Number72404231
Device Lot Number1100216286
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/27/2026
Initial Report FDA Received Date04/20/2026
Date Device Manufactured11/15/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age76 YR
Patient SexMale
-
-