| Model Number |
P8005A |
| Medical Device Problem Code |
Unintended Movement (3026)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/24/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The baxter technical support representative performed troubleshooting over the phone with the account, asking the account to replace caster.Per the baxter service manual the bed should be subject to an effective maintenance program.An annual service of the bed is advised in order to maintain its characteristics and performance.Brake casters should be checked for cuts, wear and quality of tread, etc.And replaced when necessary.Apply the brake and make sure the stretcher does not move.If there is movement, look at the brake components for wear.Apply the steer, and make sure the stretcher steers correctly.Look at the steer components for wear.Put the stretcher in neutral.Make sure all four casters rotate and roll freely.Adjust or replace components if necessary.A search of the baxter maintenance records did not show baxter performed any preventative maintenance on this bed.It is unknown if the facility performs preventative maintenance on their beds.The account will replace the bed no further action needed.
|
| |
|
Event or Problem Description
|
|
Baxter received a report that transtar stretcher had brakes not holding.There was no patient/user injury, medical intervention, symptom, or adverse event reported.
|
| |
|