| Model Number |
CI-1601-05 |
| Medical Device Problem Codes |
Migration or Expulsion of Device (1395); Adverse Event Without Identified Device or Use Problem (2993); Migration (4003)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
03/18/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The company was informed that the recipient's device was explanted.The recipient was reimplanted with another advanced bionics cochlear device.Advanced bionics is in the process of gathering more information.Once more information becomes available a supplemental report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: section b.1, b.2, h.1, h.6.Advanced bionics considers the investigation into this reportable event as closed.The recipient reportedly experienced electrode migration.A ct scan confirmed electrode migration.The recipient was successfully reactivated.The external visual inspection revealed the electrode was severed.This is believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock was verified.The condition of the electrode prevented an electrical test from being performed.The device passed the electrical and mechanical tests performed.This device was explanted for medical reasons.The device passed the tests performed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
| |
|
Search Alerts/Recalls
|