BECTON DICKINSON & CO (FRANKLIN LAKES) BD VACUTAINER® SODIUM FLUORIDE/POTASSIUM OXALATE 5MG/4MG; TUBES, VIALS, SYSTEMS, SERUM SEPARATORS, BLOOD COLLECTION
|
Back to Search Results |
|
| Catalog Number |
367921 |
| Medical Device Problem Code |
Short Fill (1575)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/23/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation summary: bd did not receive any samples or photos for investigation.However, 100 retained samples were visually inspected, and the hemogard closure assembly was found to be correctly assembled and placed on the tubes.Additionally, functional tests were performed on 10 retained samples.All tubes were within specification limits, with no issues identified.The device history records were reviewed with no issues being identified.There were no related quality notifications.All process and final inspections comply with specification requirements.This complaint has not been confirmed for the indicated failure mode: draw volume.Bd was not able to identify a root cause for the indicated failure mode.Based on an evaluation of severity and frequency it was determined that no corrective actions are required.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for the identification of emerging trends.
|
| |
|
Event or Problem Description
|
|
It was reported while using bd vacutainer® sodium fluoride/potassium oxalate 5mg/4mg, an unspecified number of tubes underfilled.No adverse impact was reported regarding the patient or user.
|
| |
|