ABBOTT MEDICAL (CRM-SYLMAR) UNIFY ASSURA CRT-D RF HV; NO MATCH Back to Search Results
Model Number CD3357-40C
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Unspecified Infection (1930); Pocket Erosion (2013)
Date of Event 04/02/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A device history record (dhr) review was performed and all required manufacturing processes and inspection steps were confirmed to be completed per the requirements.The device met specifications prior to leaving abbott manufacturing facilities.Review of the sterilization records confirmed normal sterilization cycles for the products.The product was returned and visual inspection was normal.Interrogation of the device revealed it was above elective replacement indicator (eri) when received.The cause of infection could not be traced to the device.
 
Event or Problem Description
It was reported that the patient presented to the hospital due to infection and erosion at the device site.The vegetation was noted on the right atrial lead.The pacemaker and right atrial lead were explanted and replaced.The patient was in stable condition throughout the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIFY ASSURA CRT-D RF HV
Common Device Name
NO MATCH
Manufacturer (Section D)
ABBOTT MEDICAL (CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT MEDICAL (CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
navpreet singh
15900 valley view court
sylmar, CA 91342
MDR Report Key24934961
Report Number2017865-2026-08195
Device Sequence Number17441780
Product Code NIK
UDI-Device Identifier05414734508186
UDI-Public(01)05414734508186(10)P000206629(17)260630
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P030054
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2026
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/30/2026
Device Model NumberCD3357-40C
Device Catalogue NumberCD3357-40C
Device Lot NumberP000206629
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/02/2026
Supplement Date Received by Manufacturer04/28/2026
Initial Report FDA Received Date04/21/2026
Supplement Report FDA Received Date04/28/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/24/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
LEFT VENTRICULAR LEAD.; RIGHT VENTRICULAR LEAD.
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-