| Model Number |
CD3357-40C |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Unspecified Infection (1930); Pocket Erosion (2013)
|
| Date of Event |
04/02/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A device history record (dhr) review was performed and all required manufacturing processes and inspection steps were confirmed to be completed per the requirements.The device met specifications prior to leaving abbott manufacturing facilities.Review of the sterilization records confirmed normal sterilization cycles for the products.The product was returned and visual inspection was normal.Interrogation of the device revealed it was above elective replacement indicator (eri) when received.The cause of infection could not be traced to the device.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient presented to the hospital due to infection and erosion at the device site.The vegetation was noted on the right atrial lead.The pacemaker and right atrial lead were explanted and replaced.The patient was in stable condition throughout the procedure.
|
| |
|