| Catalog Number |
SXPP1B450 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/01/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Corrected information: d 9.Device available for evaluation? additional information: d 9.Date device returned to manufacturer, d 9.Is device returned to manufacturer? h6 component code: g07002 - pending evaluation of returned device.The device upon which this medwatch is based has been received, however, the evaluation is not yet complete.Any further information derived from the evaluation will be submitted in a supplemental 3500a form.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent a robotic bilateral inguinal and umbilical hernia repair on (b)(6) 2026 and barbed suture was used.During a robotic bilateral inguinal and umbilical hernia repair, the surgeon took two passes into medtronic dextile anatomical mesh with the stratafix sxpp1b432 suture.As he was tightening the suture, it broke and he needed to remove the suture strands.He attempted to use a new stratafix two more times and the same thing occurred.He used a vloc suture to complete the inguinal hernia repairs.He then used an sxpp1b450 stratafix on the umbilical hernia repair and again the suture broke after a couple of passes.The surgeon switched to vloc to complete the surgery.He is a very experienced robotic surgeon.There was no patient consequence reported.The devices were returning.Additional information was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint#: (b)(4).Date sent to the fda: 4/21/2026.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H6 component code: g07002 - no device problem found.A manufacturing record evaluation was performed for the finished device lot, and no non-conformance's were identified.Additional information was requested, the following was obtained: our failure analysis lab received an extra product with reference to this complaint, it was received: product code: sxpp1b432.Product lot/batch: 1073he.1.Could you please clarify if extra device belongs to this complaint? 2.If not, could you please clarify if the extra received product belongs to a different complaint? if so, can you please provide the complaint number.3.If the device was not reported before, please provide more details of device malfunction.4.If the product doesn't belong to any complaint, please let us know so we can discard it or advise if the product is required to be returned.The extra devices belong to this complaint.They were sent back as they were from the same lot number and could be used for analysis purposes.A device has been received; however, it has not yet been evaluated.Any further information derived from the evaluation will be submitted in a supplemental 3500a form.
|
| |
|
Search Alerts/Recalls
|
|
|