• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. BIPAP A40; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. BIPAP A40; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE Back to Search Results
Model Number 1111169
Medical Device Problem Code Circuit Failure (1089)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/14/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The capa process identified complaints not being created for quality issues found during third party servicing.Based on a review of records impacted by the identified issue, it was determined that this record needed to be corrected and is now being submitted accordingly.The reported failure of ventilator inoperative is accommodated in the product risk management file as being linked to hazard with description loss of function/loss of primary function.A field action was initiated with record id [fca number tracked by ca?], and consisted of a field safety notice.Complaints for this failure are reviewed via the post market complaint trending and escalation processes to assess for the observed probability levels and compare them with the predicted levels in the risk file for the hazards linked to the failure.As of the date of submission for this report, there is no further indication that complaints for the failure in question has led to unacceptable increase in probability levels for the hazardous situations in scope, and therefore no further action will be pursued.Qa continues to monitor complaints for this issue per the cadence in the complaint trending procedures.
 
Event or Problem Description
A ventilator was returned to a third-party service center for service.There was no harm or injury reported.During the evaluation of the device at third-party service center, ventilator inoperative error codes were found in the ventilator's downloaded error log.The device's system board needs to be replaced to address the issue.Additionally, dust/dirt was also observed and the device was corrected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIPAP A40
Common Device Name
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 15668
7247330200
MDR Report Key24942050
Report Number2518422-2026-112449
Device Sequence Number8926357
Product Code MNT
UDI-Device Identifier00606959039476
UDI-Public(01)00606959039476
Combination Product (Y/N)N
Initial Reporter CountryPL
PMA/510(K) Number
K121623
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Y
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 04/21/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number1111169
Device Catalogue Number1111169
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/14/2025
Initial Report FDA Received Date04/21/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/04/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-1813-2024
Patient Sequence Number1
Patient SexUnknown
-
-