| Catalog Number |
222991 |
| Medical Device Problem Codes |
Break (1069); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/10/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during the unknown procedure, the fastin rc 5.0 w/ocord w/ndls anchor was handed to the specialist, and upon impact and subsequent threading, the guide on which the anchor was mounted was withdrawn; however, the attached super-sutures were found to be broken and torn, causing the anchor to detach and remain lodged within the patient's bone, rendering the titanium anchor unrecoverable.To resolve this issue, a peek anchor was provided to the physician, allowing the procedure to be completed correctly and the surgery to conclude successfully.There were no adverse effects on the patient, nor were there any alterations to the surgical timeline.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative: as of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative: investigation summary: the complaint device was not returned to depuy synthes, therefore unavailable for a physical evaluation.A manufacturing record evaluation was performed for the finished device lot number (271l222), and there was a non-conformance unrelated to the reported complaint.Based on the current available information, we cannot determine a root cause for the reported failure.If the complaint device is received in the future, we will reopen the complaint and perform the investigation as appropriate.It has been determined that no corrective and/or preventative action is proposed and there is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.
|
| |
|
Search Alerts/Recalls
|
|
|