| Catalog Number |
366593 |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
03/10/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The manufacturing location for this product is (htl).This site is an oem manufacturing site.Therefore, bd corporate headquarters in franklin lakes, nj has been listed in sections d.3.And g.1.And the franklin lakes fda registration number has been used for the manufacture report number.E.1.Initial reporter facility name: (b)(6).A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that after a blood collection procedure using bd microtainer contact activated lancet, one patient felt the sensation of a foreign body and pain.A physician examined the patient and determined the cause unknown and by this time, the pain had subsided.At a later date, the patient self excised the raised area of skin and brought the specimen to the facility.There is no report of additional medical intervention or treatment.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary: bd had not received samples or photos for investigation.Therefore, retention samples from bd inventory were evaluated by visual examination and no issues were observed relating to breaks off as all samples met specifications.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint is unable to be confirmed for the indicated failure mode breaks off.Bd was not able to identify a root cause for the indicated failure mode.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for identification of emerging trends.
|
| |
|
Event or Problem Description
|
|
It was reported that after a blood collection procedure using bd microtainer contact activated lancet, one patient felt the sensation of a foreign body and pain.A physician examined the patient and determined the cause unknown and by this time, the pain had subsided.At a later date, the patient self excised the raised area of skin and brought the specimen to the facility.There is no report of additional medical intervention or treatment.
|
| |
|
Search Alerts/Recalls
|