|
The device mentioned in this complaint has exceeded its useful life per the applicable ifu.Multiple service activities occur during the device¿s useful life according to the device¿s service manual.These service activities, in addition to any prior complaint investigations, have altered the device from its original state during manufacturing.Any issue found during manufacturing that is potentially documented in the dhr for this device does not have a causal relationship to the allegations or issues in this complaint.Therefore, the dhr for this device will not be reviewed at this time.However, the dhr files for this device are available and can be accessed upon specific request.The affected products with field complaints alleging pe-pur foam degradation have undergone the establishment of complaint codes specifically identifying foam degradation, for system one.A field correction action(2021-06-a) was initiated, and medical device recall letters were issued.A field action was initiated to replace pe-pur foam in affected products; degradation was added to the respective dfmeas and the risk management file ¿ 2203077 v27 (systemone) was updated to include hazards associated with foam degradation based on complaints analyzed through various health hazard evaluations (hhes).Risk tags ¿ degradation01, toxic02 and particulate01 reflect the safety risk of design containing the pe-pur foam.These complaints were monitored and trended in accordance with the complaint trending procedures.If the complaints were determined to meet the criteria for escalation, they were escalated for risk assessment through the post market clinical escalation process.As of the date of submission for this report, there is no indication that complaints for the failure in question have led to unacceptable levels of severity or probabilities of harm, and therefore no further action will be pursued.
|
|
A remstar auto a-flex, device was returned to a third-party service center for service as part of the sound abatement foam recall process.During the evaluation of the device, the service technician observed damaged foam.During the evaluation, there was dust contamination found in the device.Additionally, the evaluation of the device data identified unrelated error codes.These error codes are consistent with normal device operation and expected use conditions and are not considered reportable under medical device reporting requirements.No evidence was identified to suggest that these error codes contributed to or caused the reported event.The device was scrapped.
|