| Model Number |
7120Q/65 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Unspecified Infection (1930)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient's implantable cardioverter-defibrillator, the right atrial lead, the right ventricular lead and the left ventricular lead were explanted due to infection.The patient's status was not provided.
|
| |
|
Additional Manufacturer Narrative
|
|
A device history record (dhr) review was performed and all required manufacturing processes and inspection steps were confirmed to be completed per the requirements.The device met specifications prior to leaving abbott manufacturing facilities.Review of the sterilization records confirmed normal sterilization cycles for the products.The product was returned and visual inspection was normal.The cause of infection could not be traced to the device.
|
| |
|