• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LTD FISHER & PAYKEL HEALTHCARE; NASAL CANNULA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LTD FISHER & PAYKEL HEALTHCARE; NASAL CANNULA Back to Search Results
Model Number OPT944
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/23/2026
Type of Reportable Event Malfunction
Event or Problem Description
A distributor reported on behalf of a healthcare facility in china via a fisher & paykel healthcare (f&p) field representative that the opt944 optiflow + adult nasal cannula was found damaged during patient use.There were no reported patient consequences.
 
Additional Manufacturer Narrative
(b)(4).Product background: the opt944 optiflow + adult nasal cannula is a nasal cannula patient interface for delivery of humidified respiratory gases, including those who are receiving nasal high flow therapy (nhf).The cannula consists of a lightweight delivery tube connected to a rigid plastic base and soft nasal prongs (nasal interface).This allows the device to be light weight and durable.The interface is held in place by a head strap and features a head strap clip which works in tandem with the tubing clip (attaches to the patient's clothing/bedding) to support the weight of the circuit and prevent the cannula being dislodged.Optiflow + interfaces are designed for use with the airvo series humidifiers and are also compatible with the fisher & paykel healthcare (f&p) mr850 and 950 humidification system devices.Optiflow + interfaces are intended for use with spontaneously breathing patients who would benefit from receiving high flow warmed and humidified respiratory gases.Optiflow + interfaces are not intended for life support and appropriate patient monitoring must be used at all times.Method: the subject opt944 optiflow + adult nasal cannula was not returned to fisher & paykel healthcare (f&p) for evaluation.The healthcare facility reported that the subject device was discarded.F&p's investigation is based on the information and photograph provided by the healthcare facility, previous investigations of similar complaints, and f&p's knowledge of the product.Results: visual inspection of the photograph provided by the healthcare facility revealed that the head strap of the subject opt944 optiflow + adult nasal cannula was covered in adhesive tape in multiple locations.There were no patient consequences reported by the healthcare facility.Conclusion: f&p's investigation was unable to determine the exact cause of the reported damage.Based on f&p's knowledge of the product, the reported damage is likely to have been caused by the head strap being subjected to excessive force.F&p's manufacturing controls for the optiflow + tubing include inspections during production for visual defects including tears, moulding defects, discoloration and stretching or deformation.The subject opt944 optiflow + adult nasal cannula would have met the required specifications.The user instructions which accompany the opt944 optiflow + adult nasal cannula show in pictorial format the correct placement and fitting of the cannula, including ensuring the head strap clip and the tubing clip are appropriately attached.The user instructions also state the following: - "ensure head strap clip is attached, to prevent cannula from being pulled out of the nares." - "cannula can become unattached if not used with the head strap clip." - "adjust head strap to fit.Do not over-tighten." - "attach tubing clip to clothing/bedding to prevent cannula from pulling off face." - "appropriate patient monitoring must be used at all times.Failure to monitor the patient may result in loss of therapy, serious injury or death." - "failure to use the set-up described above can compromise performance and affect patient safety.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FISHER & PAYKEL HEALTHCARE
Common Device Name
NASAL CANNULA
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
diego vargas banuelos
17400 laguna canyon road
suite 300
irvine, CA 92618
9494341005
MDR Report Key24949350
Report Number9611451-2026-01406
Device Sequence Number17462408
Product Code BTT
UDI-Device Identifier09420012429117
UDI-Public(01)09420012429117(10)2104034878(11)251103
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K162553
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative,Distributor
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 04/22/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberOPT944
Device Catalogue NumberOPT944
Device Lot Number2104034878
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/23/2026
Initial Report FDA Received Date04/22/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/03/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-