Model Number CB-1000
Medical Device Problem Codes Backflow (1064); Burst Container or Vessel (1074)
Health Effect - Clinical Code Insufficient Information (4580)
Type of Reportable Event Malfunction
Event or Problem Description
The hospital reported that they have been experiencing regular problems with the blower mister: the blower mister directs part of the co2 into the bag of physiological saline solution, such that the bag ¿explodes¿ after some time.
 
Additional Manufacturer Narrative
Tw (b)(4).Event site name exceeded character limitations: (b)(6).Since the device is not available to be returned to us, a technical evaluation cannot be performed.Per our standard sop's, all events are tracked and trended to determine whether or not any trends develop.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BLOWER MISTER, 5-PACK
Common Device Name
LAVAGE, JET
Manufacturer (Section D)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ 07470
Manufacturer (Section G)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ 07470
Manufacturer Contact
arelean guzman
45 barbour pond drive
wayne, NJ 07470
MDR Report Key24950664
Report Number2242352-2026-0000296
Device Sequence Number10105941
Product Code FQH
UDI-Device Identifier00607567700741
UDI-Public(01)00607567700741
Combination Product (Y/N)N
Initial Reporter CountryBE
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 04/22/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCB-1000
Device Catalogue NumberCB-1000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/02/2026
Supplement Date Received by Manufacturer06/05/2026
Initial Report FDA Received Date04/22/2026
Supplement Report FDA Received Date06/05/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-