| Model Number |
436066 |
| Medical Device Problem Code |
Device Sensing Problem (2917)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/04/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Biotronik submits this report to comply with fda regulations 21 cfr parts 4 and 803.Biotronik has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by biotronik, or its employees that the device, biotronik, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Biotronik will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Device was explanted due to loss of sensing following cardioversion.No adverse patient events were reported.Should additional information become available, this file will be updated.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis.The analysis is thus based on the inspection of the quality documents associated with the manufacture of this particular device as well as the data returned for evaluation.The manufacturing process for this device was reviewed and all production steps were performed accordingly.There was no sign of any inconsistency during the manufacturing process.The final acceptance test proved the device functions to be as specified.The returned data were thoroughly inspected.Inspection confirmed the reported loss of signal.However, the root cause was not determinable based on the available information.It cannot be excluded that external influences, such as the reported cardioversion might have led to a component damage.Should the device itself become available for analysis, this investigation will be updated.
|
| |
|
Search Alerts/Recalls
|