| Model Number |
SIGTRSB60AMT |
| Medical Device Problem Codes |
Failure to Fire (2610); Expulsion (2933); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/25/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
According to the reporter, during distal pancreatectomy, at the time of pancreatic resection.The stapler failed to fire.Manual suturing was done to resolve the issue.It was also noted that the operation time was extended by 30 minutes.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
D10 concomitant products: sigtrsb60axt 60mm xtr thk reinforced rel, (lot # n5k1516y) egiaustnd, egiaustnd endogia ultra univ std stap, (lot # p5l1085) sigtrsb60amt 60mm med thk reinforced rel, (lot # n6a1196y) sigtrsb60axt 60mm xtr thk reinforced rel, (lot # n5k1516y) sigtrsb60axt 60mm xtr thk reinforced rel, (lot # n5k1516y) egiaustnd, egiaustnd endogia ultra univ std stap, (lot # p5l1085) medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h6 additional information received shows that there is no information to reasonably suggest that the device in this report may have caused or contributed to a death or serious injury or that the device in this report has malfunctioned.Therefore, this event does not meet the reporting requirements in 21 cfr 803.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during distal pancreatectomy, at the time of pancreatic resection, the stapler failed to fire.It was also noted that there were staples near the base of the jaw of the cartridge and the handle was idling.Manual suturing was done to resolve the issue.It was also noted that the operation time was extended by 30 minutes.There was no patient injury.
|
| |
|