| Catalog Number |
05.001.223 |
| Medical Device Problem Codes |
Mechanical Jam (2983); Complete Loss of Power (4015)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during service and evaluation, it was determined that the sagittal saw attachment device was frozen/would not move.It was noted in the service order that during pre-surgery, it was discovered that the device did not work anymore.It was not reported if there were any delays in the surgical procedure or if a spare device was available for use.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.All available information has been disclosed.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H10 additional narrative: investigation summary = > the actual device was returned for evaluation.A visual and functional assessment was performed on the device.The following steps failed: general condition check for current version check general function in running mode check the oscillation frequency with frequency meter during the service evaluation the following failures were identified: functional: frozen/will not move functional: jammed/seized - knob functional: does not function visual: component damaged.Conclusion: ¿ failure reported is confirmed.The assignable root cause was determined to be due to component failure from wear.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative: a visual and functional assessment was performed on the device.The following steps failed: general condition.Check for current version.Check general function in running mode check the oscillation frequency with frequency meter.During the service evaluation the following failures were identified: functional : frozen/will not move.Functional : jammed/seized - knob.Visual : component damaged.Conclusion: failure reported is confirmed.
|
| |
|
Search Alerts/Recalls
|