Catalog Number ES8922
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/09/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The lot number was reviewed for complaint trend, nonconforming report and capa.Devices met specification prior to release.No trends were noted for complaints and there were no nonconforming reports or capas that were confirmed to be associated.
 
Event or Problem Description
According to the available information, cylinders removed and replaced due to a tubing break "[fracture]" on the left cylinder.
 
Event or Problem Description
According to the available information, cylinders removed and replaced due to a tubing break "[fracture]" on the left cylinder.
 
Additional Manufacturer Narrative
Titan pump and cylinders 1 and 2 were received for evaluation.Abrasion was noted on both exhaust tubes and the inlet tube of the pump.No functional abnormalities were noted the pump.Abrasion was noted on both exhaust tubes of the cylinders.A separation, surrounded by abrasion, was noted on the exhaust tube of cylinder 1.This is a site of leakage.No functional abnormalities were noted with cylinder 2 based on examination of the returned product, it was concluded that while in-vivo both the exhaust tubes and inlet tube of the pump had overlapped and abraded against one another.This positioning, in combination with device usage over time, could contribute to sufficient stress to separate the exhaust tubing of cylinder 1 at the strain relief junction.A separation of this type could then allow the loss of fluid, making the device inoperable.A review of the device history record confirmed the devices from this lot met all specifications prior to release.A review of the complaint history database, nonconformances and capas revealed no trends for this lot.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TITAN INFLATABLE PENILE PROSTHESIS FAMILY
Common Device Name
PENILE PROSTHESIS, INFLATABLE
Manufacturer (Section D)
COLOPLAST A/S
1 holtedam humlebaek, dk 3050
humlebaek 3050
DA  3050
Manufacturer (Section G)
MINNEAPOLIS
Manufacturer Contact
usbes brian schmidt
1601 west river road n
minneapolis, MN 55411
MDR Report Key24955765
Report Number2125050-2026-00677
Device Sequence Number10029477
Product Code FHW
UDI-Device Identifier05708932487382
UDI-Public5708932487382
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P000006
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/25/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date10/25/2023
Device Catalogue NumberES8922
Device Lot Number6526318_ES89222400
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/20/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date04/22/2026
Supplement Report FDA Received Date06/25/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/15/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age70 YR
Patient SexUnknown
-
-