|
The customer technical support (cts) provided troubleshooting assistance.The customer replaced both reagent syringes.No further g flag errors were observed after syringes replacement.The customer was satisfied with the guidance and solutions provided and required no further assistance.Section a2, a3, a4 and a5: information was not provided.The beckman coulter internal identifier is (b)(4).
|
|
The customer reported continued g result flag errors on their au 480 clinical chemistry analyzer (pn: b96692, sn: (b)(6) and low results during fentanyl tests, which were not resolved through dilutions or routine maintenance.There was no adverse event, no harm, injury, exposure, death, or confirmed change/delay associated with this event.The customer did not provide patient data, and / or patient demographics.
|