| Model Number |
797415 |
| Medical Device Problem Code |
No Device Output (1435)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/06/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that an intrasight mobile system was to be used for a stent placement procedure.However, during use, the system was unable to recognize the ivus-pim and displayed a "connect pim" error, even when connected.The procedure was stopped since ivus was needed for proper stent sizing, and rescheduled for a later date.No patient injury reported.This product problem is being reported because the system could not be used; resulting in a potential for harm due to the delay of the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
This case was reviewed and investigated according to the manufacturer¿s policy.Blocks a2 & a3b-a6: not available from facility.Blocks b6 & b7: not available from facility.Block c: not applicable for this system.Blocks d4, d6 & d7: not applicable for this system.Blocks h3 & h6: philips field service engineer determined the pim cable required replacement.A pim cable was sent to the facility for replacement, the old pim cable was discarded; no further issues have been reported.Blocks h7, h9 & h10: do not apply to this submission.Submission of this report does not, in itself, represent a conclusion by the manufacturer and/or authorized representative or the national competent authority that the content of this report is complete or accurate, that the medical device(s) listed failed in any manner and/or that the medical device(s) caused or contributed to an alleged death or deterioration in the state of the health of any person.
|
| |
|
Search Alerts/Recalls
|
|
|