• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DENTSPLY IMPLANTS MANUFACTURING GMBH V-TWIST DRILL EV Ø4.3, 6-17MM; ACCESSORIES, IMPLANT, DENTAL, ENDOSSEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DENTSPLY IMPLANTS MANUFACTURING GMBH V-TWIST DRILL EV Ø4.3, 6-17MM; ACCESSORIES, IMPLANT, DENTAL, ENDOSSEOUS Back to Search Results
Catalog Number 25783
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/08/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
There has been a previous report received where this malfunction resulted in a serious injury.Therefore, is must be presumed that recurrence of this malfunction could possibly cause or contribute to a serious injury or require medical or surgical intervention to preclude such.As such, this event is reportable per 21cfr part 803.Product return is requested and product will be evaluated after receipt.In case any new or additional information will be gained from this investigation a follow-up report will be sent.Trend is tracked and monitored.
 
Event or Problem Description
It was reported that after the first use, it fell out of the adapter/handpiece during use.No injury resulted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V-TWIST DRILL EV Ø4.3, 6-17MM
Common Device Name
ACCESSORIES, IMPLANT, DENTAL, ENDOSSEOUS
Manufacturer (Section D)
DENTSPLY IMPLANTS MANUFACTURING GMBH
rodenbacher chaussee 4
hanau hessen 63457
GM  63457
Manufacturer (Section G)
DENTSPLY IMPLANTS MANUFACTURING GMBH
rodenbacher chaussee 4
hanau hessen 63457
GM   63457
Manufacturer Contact
dan eagar
221 west philadelphia st.
york, PA 17401
7178457511
MDR Report Key24957360
Report Number3013111692-2026-13804
Device Sequence Number20149913
Product Code NDP
UDI-Device Identifier07392532146428
UDI-Public07392532146428
Combination Product (Y/N)N
Initial Reporter CountryPL
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 04/22/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number25783
Device Lot Number540989
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/10/2026
Is the Reporter a Health Professional? Yes
Distributor Facility Aware Date04/10/2026
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/10/2026
Initial Report FDA Received Date04/22/2026
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NOT PROVIDED
Patient SexUnknown
-
-