| Model Number |
1701 |
| Medical Device Problem Codes |
Fracture (1260); Unexpected Therapeutic Results (1631); Communication or Transmission Problem (2896)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/26/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that there were 6 cases of lead breakage/fracture, in all cases at the identical side of the lead, directly located at the transition from the rigid section to its flexible section.These all occurred during trial stages after the lead had been placed in the patient.In most cases, the fracture resulted in loss of isolation and exposure of the cable core/strand with short circuit, loss of connectivity to the patient's remote, loss of device function, and termination of the testing phase due to the removal of the lead.The physician is having to consider switching back to the previous trial device they've used for the time being because of the issue that occurred in the most recent 6 cases until the issue is resolved.
|
| |
|
Additional Manufacturer Narrative
|
|
Block d2b: product code - additional product code qon.Block g4: premarket/510(k) # - additional premarket/510(k) p190006.
|
| |
|
Event or Problem Description
|
|
It was reported that the during the trial stage the percutaneous extension lead (pne) fractured at the transition from the rigid section to the flexible section.The fractured resulted in loss of isolation, exposure of the cable core or strand with short circuit, loss of connectivity to the patient's remote, loss of device function and termination of the testing phase due to removal of the lead.No further patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Block d2b:
product code - additional product code qon
block g4:
premarket / 510(k) # - additional premarket/510(k) p190006
block h11:
corrections:
block b1: adverse event or product problem
block b2: outcomes attributed to adverse events
block b5: describe event or problem
block d3: manufacturer email
block h6: device codes
this event does not report any product malfunction, nor is there any allegation of patient/user harm.Upon further review by the manufacturer, this was determined to not meet the criteria for a reportable event.
|
| |
|