• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: THE ANSPACH EFFORT LLC 5 CM HEAVY DUTY SHORT ATTACHMENT; MOTOR, DRILL, ELECTRIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

THE ANSPACH EFFORT LLC 5 CM HEAVY DUTY SHORT ATTACHMENT; MOTOR, DRILL, ELECTRIC Back to Search Results
Catalog Number SHORT-HD-G1
Medical Device Problem Code Excessive Heating (4030)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during service and evaluation, it was determined that the attachment device had heat.It was further determined that the device failed pretest for temperature verification.This event did not occur during surgery.There were no reports of injuries, medical intervention, or prolonged hospitalization.The exact date of the event was not reported; however, it was reported that the event occurred in 2026.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H11: additional narrative: device evaluation: the actual device was returned without an alleged deficiency.During routine service and repair, it was determined that the device had heat.Therefore, the reported condition was confirmed.The assignable root cause of this condition was determined to be traced to component failure due to wear.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5 CM HEAVY DUTY SHORT ATTACHMENT
Common Device Name
MOTOR, DRILL, ELECTRIC
Manufacturer (Section D)
THE ANSPACH EFFORT LLC
4500 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
elba bello
4500 riverside drive
palm beach gardens, FL 33410
9083863534
MDR Report Key24958700
Report Number1045834-2026-00094
Device Sequence Number13341117
Product Code HBC
UDI-Device Identifier00845384016304
UDI-Public(01)00845384016304
Combination Product (Y/N)N
Initial Reporter CountryNL
PMA/510(K) Number
K133604
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberSHORT-HD-G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/30/2026
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/14/2026
Initial Report FDA Received Date04/22/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/06/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-