| Catalog Number |
50000000L |
| Medical Device Problem Codes |
Restricted Flow rate (1248); Failure to Pump (1502)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/17/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that per wo evaluation, arctic sun device control panel booted to red screen during assessment.Power cycles 5 times and the control panel booted up normally.During repair discovered no flow reading /will be replaced flow meter to correct.
|
| |
|