Date of event is unknown.Additional information will be provided if available.The device was returned to olympus for inspection.A root cause could not be identified.Based on the results of the investigation, it is likely deposition of foreign material in nozzle resulted in clogged nozzle.However, the most probable cause of occurrence of foreign material could not be established.However, use of products such as simethicone, lubricants that contain silicone can cause deposits of white foreign material and thus are not recommended for use.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|