ZIMVIE US CORP LLC T3® NON-PLATFORM SWITCHED TAPERED IMPLANT 4 X 10MM; DENTAL IMPLANT Back to Search Results
Catalog Number BOST410
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Unspecified Infection (1930); Osteolysis (2377)
Date of Event 03/03/2026
Type of Reportable Event Serious Injury
Event or Problem Description
The doctor reports that the dental implants located in dental positions 25 and 27 failed due to peri-implantitis.The doctor reports in the per that the procedure was not concluded by placing another implant.
 
Additional Manufacturer Narrative
Zimvie complaint number: (b)(4).A4: patient weight: unknown / not provided.D10: concomitant medical product: bost585, t3® non-platform switched tapered implant 5 x 8.5mm, lot: 2120024484.A summary investigation has been completed for peri-implantitis events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for these events have not identified any signals indicating potential non-conformance's affecting the manufacturing and sterilization processes.Furthermore, the probability of a manufacturing or design defect that might lead to peri-implantitis occurring and escaping the available detections has been assessed and found remote and almost nonexistent.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
T3® NON-PLATFORM SWITCHED TAPERED IMPLANT 4 X 10MM
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
ZIMVIE US CORP LLC
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
ZIMVIE US CORP LLC
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
mark mashburn
4555 riverside drive
palm beach gardens, FL 33410
5619713230
MDR Report Key24964103
Report Number0001038806-2026-02220
Device Sequence Number4849774
Product Code DZE
UDI-Device Identifier00844868028925
UDI-Public(01)00844868028925(17)300122(10)2120025804
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K122300
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 04/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Catalogue NumberBOST410
Device Lot Number2120025804
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/09/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date04/23/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/23/2025
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DENTAL IMPLANT SEE H10 NARRATIVE.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age64 YR
Patient SexMale
Patient EthnicityNon Hispanic
-
-