|
Hamilton medical reference number: b4 this report also contains the investigation outcome.Hamilton medical ag received the logfiles of the device for analysis.Upon review of the logfiles, the issue reported by the user could be confirmed.The device alarmed with technical fault - tf 5509 and tf 9907 on the day of the event, b6.It is important to emphasize that no patient was involved at the time the alarm occurred.Please see below extract from the logfiles.B6 17:20:02 tf : 9907 tech fault 9907 b6 17:20:01 tf : 2803 15295 tech fault 2803 b6 17:20:01 tf : 2801 500 tech fault 2801 b6 17:20:01 tf : 2804 1 tech fault 2804 b6 17:20:01 tf : 2402 6 tech fault 2402 b6 17:20:01 tf : 2414 tech fault 2414 b6 17:20:01 audio paused off special 517 b6 17:20:01 audio paused off special 517 b6 17:20:01 ambient irreversible setting 233 b6 17:20:01 tf : 5509 tech fault 5509 according to the service manual for hamilton-g5, tf5509 indicates that the error servo signal remains active for 5 seconds, indicating that the servo (inspiratory) valve is not functioning correctly.Further on, the tf 9907 indicates faulty communication between integrated panel processor (ipp) (panel) and ventilation unit processor vup (vu).Therefore, to address the issue, the inspiratory valve assembly was replaced.Following the replacement, full calibration and functional testing were performed, confirming that the device operates in accordance with its intended performance specifications.Hamilton medical considers this case closed.
|