It was reported that the tracheal tube cuff deflated immediately upon re-intubation, following a similar failure and requiring the patient a total of four times, three more times following this incident.According to the reporter, during a stomatological procedure, a size 6.0 tracheal tube was used for left nasal intubation via the left nostril with the assistance of a video laryngoscope.Passage through the nasal turbinates was achieved without difficulty.Intubation was described as easy, with good visualization of the vocal cords, and magill forceps were not utilized.The reporter stated that during attempted cuff inflation, inflation failure was observed, as the cuff progressively deflated during inflation.Per the anesthesiologist, no issue was identified with the cuff valve.The initial tube was replaced with a second tube of the same type and size; however, the same cuff deflation issue was observed.A third tube change was then performed using a reinforced (armored) size 5.5 tube.Intubation was successful; however, the cuff deflation issue persisted.A fourth tube, a reinforced tube obtained from the equipment reserve, was subsequently used, and the reported issue did not recur.
|