MAUDE Adverse Event Report: ICU MEDICAL HEALTHCARE MANUFACTURING S.A. DE C.V. PORTEX TRACHEAL TUBE CLEAR MURPHEY EYESOFT SEAL; TUBE, TRACHEAL (W/WO CONNECTOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ICU MEDICAL HEALTHCARE MANUFACTURING S.A. DE C.V. PORTEX TRACHEAL TUBE CLEAR MURPHEY EYESOFT SEAL; TUBE, TRACHEAL (W/WO CONNECTOR) Back to Search Results
Catalog Number 100/199/080
Medical Device Problem Code Gas/Air Leak (2946)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/22/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
One used device was returned for investigation.The device was visually inspected and there were no damages or anomalies observed.During the functional testing, the trach tube was then submerged underwater to facilitate leak detection.Initially, there was no leakage observed.After five minutes of the cuff remaining inflated, the cuff weld channel grew until a leak occurred.The leak originated from the trach tube cuff weld (proximal).This was observed for both samples.The complaint of leakage can be confirmed.The probable cause is due to a cuff welding error during manufacturing.A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.
 
Event or Problem Description
It was stated that the intensive care unit reportedly encountered a leak at the junction with the balloon.There was no reported patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX TRACHEAL TUBE CLEAR MURPHEY EYESOFT SEAL
Common Device Name
TUBE, TRACHEAL (W/WO CONNECTOR)
Manufacturer (Section D)
ICU MEDICAL HEALTHCARE MANUFACTURING S.A. DE C.V.
ave calidad no. 4, parque
tijuana 22425
MX  22425
Manufacturer Contact
reed covert
6000 nathan lane north
minneapolis, MN 55442
2247062300
MDR Report Key24965843
Report Number3012307300-2026-04114
Device Sequence Number16066133
Product Code BTR
UDI-Device Identifier15019315022340
UDI-Public(01)15019315022340(17)300108(10)6074953
Combination Product (Y/N)N
Initial Reporter CountryBE
PMA/510(K) Number
UNKNOWN
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 04/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number100/199/080
Device Lot Number6074953
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/17/2026
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/31/2026
Initial Report FDA Received Date04/23/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/23/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age66 YR
Patient SexUnknown
Patient Weight73 KG
-
-