| Catalog Number |
1460000000 |
| Medical Device Problem Code |
Mechanical Jam (2983)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
This record is a consolidation of records summarized as a part of the fda voluntary malfunction summary reporting program.Events occurred between january 1- march 31, 2026.This report summarizes 12 malfunction events(s), where it was reported the devices experienced brakes cannot be engaged; false engagement of brakes.No adverse consequence or clinically relevant delay in treatment was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This record is a consolidation of records summarized as part of the fda voluntary malfunction summary reporting program.8 devices were functionally/visually inspected in the field.The devices were repaired and returned to use.Evaluation results findings (components/results) 1 device: actuator / mechanical problem identified 3 devices: part/component/sub-assembly term not applicable / mechanical problem identified 2 devices: actuator; gears / mechanical problem identified 1 device: chassis/frame / deformation problem 1 device: foot pedal / deformation problem 1 device was not evaluated, as a probable cause for the issue was identified during communication between the customer and stryker and no further assistance was requested by the customer.Evaluation results findings (components/results) 1 device: actuator / mechanical problem identified 3 devices are pending evaluation.Evaluation results findings (components/results) 3 devices: insufficient information / results pending completion of investigation there was no remedial action taken.This device is not labeled for single use.
|
| |
|
Search Alerts/Recalls
|