This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative: a visual and functional assessment was performed on the device.The following steps failed: section 10: general condition section 70: check for sticky trigger section 80: check function of device.During the service evaluation the following failures were identified: functional : moving parts do not move smoothly - trigger functional : device stops by itself visual : contact damage.Conclusion: failure reported is not confirmed root cause: component failure from wear.
|