Additional information: h6.Correction: it was determined after receiving further clarification, this event for this product was not a reportable event per cfr 21, part 803.This supplemental is being filed to identify an mdr was not required for this event.The reported infection and resulting remake request were investigated.Review of sterilization records, environmental monitoring data, and packaging controls identified no deviations.Medical assessment concluded that the device did not cause or contribute to the reported event.Based on the investigation, there is no indication for an incorrectly working product or any design, material, or manufacturing related issue.Therefore, no corrective and/or preventive actions are deemed necessary at this time.
|