• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SURGICAL, INC. Z.S.G.M. CUTTER 1.5:1 RATIO ( CAUTION: SHARP ); EXPANDER, SURGICAL, SKIN GRAFT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL, INC. Z.S.G.M. CUTTER 1.5:1 RATIO ( CAUTION: SHARP ); EXPANDER, SURGICAL, SKIN GRAFT Back to Search Results
Catalog Number 00770301500
Medical Device Problem Code Failure to Cut (2587)
Health Effect - Clinical Code Insufficient Information (4580)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This event is recorded with zimmer biomet under (b)(4).This medwatch is being filed as an initial / final report based on information discovered during the device evaluation.Review of the most recent repair record determined the cutter failed the sample mesh test due to damaged; however, the repair was not completed because cutters are not repairable devices.The cutter was returned as is.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.The root cause of the reported issue is attributed to component failure as the device exhibits wear and tear from repeated use.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.G2: foreign: canada.
 
Event or Problem Description
It was reported that the unit was not meshing.There is no information provided regarding the timing of the event or if there was any patient involvement/harm that may have occurred.During device evaluation it was discovered that the cutter failed the sample mesh test due to damage.Due diligence is complete.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Z.S.G.M. CUTTER 1.5:1 RATIO ( CAUTION: SHARP )
Common Device Name
EXPANDER, SURGICAL, SKIN GRAFT
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
jennifer rapsavage
56 e. bell drive
warsaw, IN 46582
5745260384
MDR Report Key24967508
Report Number0001526350-2026-00490
Device Sequence Number10496223
Product Code FZW
UDI-Device Identifier00889024375444
UDI-Public(01)00889024375444(11)200710(10)64803448
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
PREAMENDMENT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 04/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number00770301500
Device Lot Number64803448
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/11/2025
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/09/2026
Initial Report FDA Received Date04/23/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/10/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-