Block e1 the initial reporter's title is clinical research coordinator.Block h6 device code a0401 is being used to capture the reportable event of cinch wire break.Device evaluation: block h11.The suturing cinch was not returned for evaluation, and no media was provided for analysis.The reported event of cinch wire break could not be confirmed.It was confirmed this device met manufacturing specification prior to distribution and there were no manufacturing deviations which could have contributed to the reported event.Based on all gathered information, there is not enough evidence to determine whether the cinch wire break was due to the physician's technique during the procedure or was related to a device malfunction.Without proper evaluation of the device, it remains unknown the most probable causes that contributed to the event.Also, the evidence from the product record review did not identify a potential product quality issue or new patient harm.Based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of cause not established.
|