There was an allegation of questionable results from coaguchek xs meter, serial number (b)(6).The meter result was 5.5 inr.Within 5 minutes, the meter result was 7.7 inr.The laboratory result using an unknown method was 2.8 inr.The exact time of the laboratory test was not provided.The patient¿s therapeutic range was not provided.
|
|
The product has been requested for investigation.The product has not been received at this time.If the product is returned in the future, a follow up report will be submitted.On a regular basis, coaguchek strips of lots currently valid in the market are tested as part of routine retention testing and results have passed the internal inspection.Per product labeling, "coaguchek uses human recombinant thromboplastin.Therefore, the comparability to other human recombinant thromboplastins is best, whereas higher deviations can occur with other thromboplastins.However, those higher differences between thromboplastins of different (rabbit, bovine based) origin are not a coaguchek specific issue.Similar differences can be observed when a human recombinant thromboplastin-based laboratory method is compared against several other (rabbit, bovine-based) laboratory methods.".
|