BOSTON SCIENTIFIC CORPORATION VISIONIST? CRT-P; Cardiac resynchronization therapy pacemaker Back to Search Results
Model Number U226
Medical Device Problem Code Premature Discharge of Battery (1057)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/14/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This product investigation is still in progress.This report will be updated when product investigation is complete.
 
Event or Problem Description
It was reported that this cardiac resynchronization therapy pacemaker (crt-p) recorded a code 1003, indicating the voltage is too low for the projected remaining capacity.A request was made to have data from this device analyzed.Data analysis identified potential high impedance within the battery.Continued monitoring was recommended.No adverse patient effects were reported.This device remains in service at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISIONIST? CRT-P
Common Device Name
Cardiac resynchronization therapy pacemaker
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
Boston Scientific Corporation
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
+1(651)582-6168
MDR Report Key24969059
Report Number2124215-2026-22121
Device Sequence Number10580862
Product Code NKE
UDI-Device Identifier00802526559457
UDI-Public(01)00802526559457(17)220415
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
P030005
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Reporter Type Manufacturer
Report Source Company Representative
Remedial Action Notification
Type of Report Initial,Followup
Report Date (Section B) 08/06/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date04/15/2022
Device Model NumberU226
Device Catalogue NumberU226
Device Lot Number701440
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/10/2026
Supplement Date Received by Manufacturer08/03/2026
Initial Report FDA Received Date04/23/2026
Supplement Report FDA Received Date08/07/2026
Date Device Manufactured09/14/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age79 YR
Patient SexFemale
-
-