• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ORTHO-CLINICAL DIAGNOSTICS, INC. VITROS 5600 INTEGRATED SYSTEM; CHEMISTRY ANALYZER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ORTHO-CLINICAL DIAGNOSTICS, INC. VITROS 5600 INTEGRATED SYSTEM; CHEMISTRY ANALYZER Back to Search Results
Catalog Number 6802413
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/23/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation determined that the sample id (sid) for two patient samples were mis-associated with the incorrect patient names and results when processed on a vitros 5600 integrated system.The assignable cause of the event was determined to be user error.The analyzer was operating as intended.The customer reused an sid without ensuring the previously programmed sid was processed to completion or deleted prior to reuse.The technical solution center reviewed information contained in the "sample id reuse" section of the 5600 integrated system reference guide.This information describes how to properly manage sids that were previously used and not processed to completion or deleted.If a sample is not processed to completion, the sample programming and associated demographics data are retained by the analyzer in a "pending" state, as per the design of the software.Reusing a sample id on a different sample without completion of the original request or the deletion of the pending testing will cause the original information to be carried forward.No malfunction occurred.
 
Event or Problem Description
A customer contacted the ortho clinical diagnostics (ortho) technical solution center (tsc) to report that the sample id (sid) for two different patient samples were mis-associated with the incorrect patients name and results when processed on a vitros 5600 integrated system.This event meets the criteria for a reportable event, as test results could have been misreported out of the laboratory leading to inappropriate intervention with the potential for serious injury to the patient.The customer identified the discrepancy, and no mis-associated patient results were reported out of the laboratory.There was no allegation of patient harm as a result of this event.This report corresponds to ortho clinical diagnostics inc (ortho) complaint number (b)(4) and reportability assessment (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VITROS 5600 INTEGRATED SYSTEM
Common Device Name
CHEMISTRY ANALYZER
Manufacturer (Section D)
ORTHO-CLINICAL DIAGNOSTICS, INC.
100 indigo creek drive
rochester NY 14626
Manufacturer (Section G)
ORTHO-CLINICAL DIAGNOSTICS, INC.
100 indigo creek drive
rochester NY 14626
Manufacturer Contact
nancy hewitson
100 indigo creek drive
rochester, NY 14626
5854534165
MDR Report Key24971021
Report Number1319681-2026-00027
Device Sequence Number20072271
Product Code JJE
UDI-Device Identifier10758750002740
UDI-Public(01)10758750002740
Combination Product (Y/N)N
Initial Reporter CountryFR
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 04/22/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number6802413
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/26/2026
Initial Report FDA Received Date04/23/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/18/2010
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-