| Model Number |
FX2100X15B |
| Medical Device Problem Codes |
Electrical /Electronic Property Problem (1198); Mechanical Problem (1384); Device Sensing Problem (2917)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/26/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The manufacturer received information alleging a trilogy evo o2 ventilator experienced a ventilator inoperative condition.There was no serious patient harm or injury that occurred or was reported.During evaluation of the device at the third party service center, the ventilator inoperative condition was confirmed.The service technician states that the machine flow sensor and fio2 tubing were replaced to resolve the issue.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported ventilator inoperative condition is accommodated in the product risk management file as being linked to hazard with description loss of function/loss of primary function.Complaints for this failure are reviewed via the post market complaint trending and escalation processes to assess for the observed probability levels and compare them with the predicted levels in the risk file for the hazards linked to the failure.As of the date of submission for this report, there is no indication that complaints for the failure in question has led to unacceptable increase in probability levels for the hazardous situations in scope, and therefore no further action will be pursued.Qa continues to monitor complaints for this issue per the cadence in the complaint trending procedures.
|
| |
|