• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SàRL INTRAFIX; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SàRL INTRAFIX; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE Back to Search Results
Catalog Number 254807
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/14/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during an arthroscopic anterior cruciate ligament repair procedure that the intrafx advbr scw8x30 w/smshth device sheath broke during implantation.Another device was used to complete the procedure.There was no delay in the procedure reported.There were no adverse patient consequences reported.All available information has been disclosed.
 
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative: as of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated, and no conclusion can be drawn at this time.The product was not returned to depuy synthes, however a photo was provided for evaluation.Visual inspection of the returned photo found that the sheath was broken and had foreign matter, presumably biological matter.In addition, the screw had signs of usage.No other anomalies were noted.The overall complaint was confirmed as the observed condition of the intrafx advbr scw8x30 w/smshth would have contributed to the complained device issue.Based on the investigation findings, the possible root cause is traced to axial misalignment during the sheath placement with the sheath inserter tool and/or an incorrect inserter tool size.As per ifu-114007: 1) mallet sheath into tunnel until sheath tab is flush on cortical bone.Remove the sheath inserter from the tunnel leaving the guidewire and sheath in place.2) place the appropriate size screw onto the hex driver.Refer to table 1, implant sizing guideline.3) pull to desired tension and insert screw over guidewire into the sheath until it is flush with the cortex of bone.Properly handling and¿attention to the approved use of the device diminishes the risk of failure.It has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Document specification review: ifu- 114007 device history lot pn: 254807 lot number: 107x73 there was no non-conformance regarding this lot manufacturing date: 22/apr/2025 expiry date: 31/mar/2028 device history batch: null device history review: pn: 254807 lot number: 107x73 there was no non-conformance regarding this lot manufacturing date: 22/apr/2025 expiry date: 31/mar/2028.E1: the reporter¿s complete facility address was not provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTRAFIX
Common Device Name
FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL
chemin-blanc 38
le locle 02400
SZ  02400
Manufacturer Contact
elba bello
325 paramount drive
raynham, MA 02767
9083863534
MDR Report Key24971424
Report Number1221934-2026-01642
Device Sequence Number10580924
Product Code MAI
UDI-Device Identifier10886705026739
UDI-Public(01)10886705026739(10)107X73(17)280331
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number254807
Device Lot Number107X73
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/15/2026
Supplement Date Received by Manufacturer06/10/2026
Initial Report FDA Received Date04/23/2026
Supplement Report FDA Received Date06/15/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/22/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-