• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. COATED VICRYL; SUTURE, ABSORBABLE, SYNTHETIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. COATED VICRYL; SUTURE, ABSORBABLE, SYNTHETIC Back to Search Results
Catalog Number VCP304H
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/27/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4) this report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Additional information was requested, and the following was obtained: how many vcp304h devices demonstrated the alleged deficiency ("the needle detached from the thread")? did this event contribute to any patient adverse event? if yes, please explain.Attached image (img_0001) shows a non-eth suture product, could you please confirm whether the mentioned product is from ethicon (a johnson and johnson company)? if different, please explain.Please confirm if this component belongs to the reported event.If not, would you kindly create a product complaint for each event and provide the corresponding reference number(s)? please confirm if a deficiency was experienced with the mentioned component.If yes, please explain.Please provide the product code and lot number.Please perform and document the follow up attempt for product return.Please document the shipment tracking number in notes or rmao sections.Please provide the title of external person providing answers to follow-up (e.G.Nurse, surgeon, risk manager).The following information was received: 5 blisters no injured patients and also, i have attached a correct photo.A photo upon which this medwatch is based has been received, however, the photo evaluation is not yet complete.Any further information derived from the evaluation will be submitted in a supplemental 3500a form.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent.It was reported on note a-15290565 "¿no injured patients." did the reported event occur during one or multiple procedures? if this occurred in multiple procedures, please confirm the number of patients affected by the alleged deficiency.Have any of these events been previously reported to ethicon? if yes, provide the respective reference number(s).Please create a new pc file for each patient/event.-how many devices demonstrated the reported alleged deficiency? h6 component code: g07002 no device problem found h3 video analysis: this is an analysis for a photo submitted to ethicon for evaluation.During the visual analysis, the following was observed: the video shows the user dispensing the suture and holding the needle with surgical forceps.When force is applied to the suture, it breaks.The video is not clear enough to determine how much force was applied to the suture.Based on the video review, the event reported is confirmed, however no conclusion or root cause could be determined.Hands on device analysis may provide the additional evidence necessary to confirm the root cause of the reported event.Because the device was not returned our evaluation is limited.We value the opportunity to fully analyze the device upon its return.As part of ethicon quality process all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post market surveillance.
 
Event or Problem Description
It was reported that a patient underwent an unknown procedure on (b)(6) 2026 and suture was used.During the suturing of the patient¿s soft tissues, the needle detached from the thread.A new blister was opened.The wound was closed using another suture material.There were no adverse patient consequences reported.Additional information was requested.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COATED VICRYL
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-GMBH
robert-koch strasse 1
p.o. box 1409
norderstedt
GM  
Manufacturer Contact
orla o'mahony
1000 route 202
raritan, NJ 08869
3035526892
MDR Report Key24971438
Report Number2210968-2026-04434
Device Sequence Number10455637
Product Code GAM
UDI-Device Identifier10705031052116
UDI-Public(01)10705031052116(17)270430(10)UEBBMCW0
Combination Product (Y/N)Y
Initial Reporter CountryRS
PMA/510(K) Number
K132580
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/12/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberVCP304H
Device Lot NumberUEBBMCW0
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/22/2026
Supplement Date Received by Manufacturer04/28/2026
Initial Report FDA Received Date04/23/2026
Supplement Report FDA Received Date05/12/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/07/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-