Svd is a logical and expected consequence of the chemical, mechanical, and immunological processes associated with bhv implantation.The common endpoint of all these factors is calcification and degradation.Svd is an acquired condition, intrinsic to bioprosthetic valves, characterized by deterioration of the leaflets or supporting structures, leading to thickening, calcification, tearing, or disruption of the prosthetic valve materials.These changes result in valvular dysfunction manifesting as stenosis and/or regurgitation with a consequent drop in hemodynamic efficacy of the valve.Prior investigations and extensive literature review have shown that svd is predominantly related to patient factors and implant duration rather than to manufacturing issues.Events related to device design, manufacturing, or use error tend to manifest at an earlier implant duration.There is no evidence to suggest a manufacturing defect caused or contributed to the reported event.Based on the information available, the most likely root cause is patient factors, including years, hld, and cad.Edwards will continue to review and monitor all events.Trends are monitored on a the investigation is still in progress; therefore, a conclusion has yet to be established.
|
|
It was reported and learned through investigation.A patient with a 27mm 11500a aortic valve implanted five (5) years, eight (8) months, underwent valve-in-valve procedure due to stenosis and calcification.Patient presented with shortness of breath.Tavr was successfully performed with a 29mm 9755rsl transcatheter valve.Patient was stable at the end the of the procedure.
|